Download Cut Sheet
The MSA Bacharach LS-25 Series Leak Standard provides an NIST - traceable reference leak source for instrument functional testing or adjustment. The unit is factory calibrated and shipped fully charged with the selected refrigerant, sufficient to provide years of service. Annual re-calibration is suggested.
With the LS-25 external leak source, the H25-IR PRO can be calibrated to the gas and leak rate desired.
Product Attributes:
- Calibration Gas: R290
- Leak Rate: 3-10 g/yr
- Accuracy: ± 15% of
- Temperature: 20 ± 2°C (68 ± 4°F)
- Humidity: 45% ± 10% RH, non-condensing
- Air Pressure: 1 Standard Atmosphere ± 10%
- Orientation: Bench Top
- Enclosure Dimensions: 6” W × 6” H × 12” D
- Weight: 3 lbs
- Power: No power supply required
Part Number: 3015-5649